Thermo Scientific* Aquakem 250 Photometric Analyzer is a discrete, random access analyzer for a wide variety of water and environmental tests.
Product Detail
The Aquakem 250 is an easy to use chemistry analyzer capable of random access or batch analysis and the use of multiple chemistries. Types of matrices analyzed include clean water, wastewater, soil/sludge digests, effluents and saline. The Aquakem system can be custom-programmed to meet individual needs in each laboratory. All method protocols are based to those recommended in the standard methods for the examination of water and wastewater in different countries and those published in the U.K. Blue Book.
Provides convenient automated photometric analysis of water and environmental samples
Analytical rate of up to 250 tests per hour
Automatic start-up and shut down protocols for true walk-away analysis
No reagent priming is required, enabling results from the first sample to be available within minutes of starting the analysis
Ideal for: Ammonia, Alkalinity, Aluminium, Boron, Calcium, Chloride, Cyanide (post distillation), Hexachrome, Ferrous ion, Magnesium, Nitrate, Nitrite, Phosphate (ortho and total post digestion), Silica, Sulphate, Total Hardness, and TKN post digestion.
Colorimetric end-point, linear and non-linear reactions, turbidimetric and bichromatic with or without sample blanking.
Reaction Vessel :
Disposable multicell cuvetes. Continuous loading of cuvetes with capacity for 600 tests.
Power Requirements :
100–240V +/-10%, 50–60Hz +/-5%, less than 400W
Samples :
Suitable for raw, potable, wastewater, saline and boiler water. Max on-board capacity 6 segments each with 14 positions for samples. Continuous access without interrupting test processing.
Sample Dilution :
Automatic sample dilution of overrange tests with automatic re-run. Manual dilution by operator if required, corrected result identified.
Weight :
142kg
Distilled Water Consumption :
Distilled water consumption typically <1L/hr. No plumbing required.
Reagents :
On-board 35 pos. in the cooled reagent disk. All reagents are automatically identified using bar code labels. Real time reagent status clearly displayed. Up to four reagent additions/tests possible. Continuous access to reagents without interrupting test processing.
Analytical Approach :
Discrete, selective, with either random access or batch analysis.